    The Practice of Chest Radiography Using Different Digital Imaging Systems: Dose and Image Quality

    (ندگان)پدیدآور
    Moey, Soo-FoonShazli, ZubirSayed, Inayatullah Shah
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    365.3کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: The study was undertaken to evaluate the practice of chest radiography using different digital imaging systems and its influence on dose and image quality.  Materials and Methods: The study was carried out in two hospitals from March 2016 to June 2016. Sixty ambulatory patients aged 21 to 60 years who were able to cooperate without difficulty and weighed between 60 to 80 kg were selected randomly. The active matrix flat panel imagers technology was employed in the direct radiography (DR) system for Hospital A, whilst Hospital B used the single read out computed radiography (CR) system. The dose area product (DAP) meter was utilized in measuring the entrance surface air kerma. The chest radiographs were evaluated by two radiologists.  Results: The mean entrance surface doses (ESDs) for posteroanterior chest in Hospital A (0.098 mGy) was lower than that obtained in Hospital B (0.161 mGy). However, the ESDs at both centres were lower than the recommended value by the International Atomic Energy Agency (IAEA; 0.3 mGy). The quality of the images for chest radiography in both hospitals was adequate to make a diagnosis with ESDs and effective doses lower than those recommended by IAEA and United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR).  Conclusion: The study serves to highlight the practice of chest radiography with two different systems and its influence on image quality and dose. It can be concluded that there were significant differences in image quality and radiation dose for chest radiography practice using CR and DR.
    کلید واژگان
    Digital Radiography
    Image Quality
    Radiation Dosage
    Thoracic Radiography
    Medical Physics
    Physics of Diagnostics Radiology

    شماره نشریه
    2
    تاریخ نشر
    2018-04-01
    1397-01-12
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    International Islamic University Malaysia
    Department of Diagnostic Imaging and Radiotherapy, Kulliyah of Allied Health Sciences, International Islamic University Malaysia
    International Islamic University Malaysia

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2017.25424.1259
    http://ijmp.mums.ac.ir/article_9583.html
    https://iranjournals.nlai.ir/handle/123456789/324780

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب