    Evaluation of the Quality Control Program for Diagnostic Radiography and Fluoroscopy Devices in Syria during 2005-2013

    (ندگان)پدیدآور
    Kharita, M. H.Wannus, K. M.Khedr, M. S.
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.016 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: Extensive use of diagnostic radiology is the largest contributor to total population radiation doses. Thus, appropriate equipment and safe practice are necessary for good-quality images with optimal doses. This study aimed to perform quality control (QC) audit for radiography and fluoroscopy devices owned by private sector in Syria (2005-2013) to verify compliance of performance of X-ray machines with the regulatory requirements stipulated by the national regulatory body. Materials and Methods: In this study, QC audit included 487 X-ray diagnostic machines, (363 radiography and 124 fluoroscopy devices), installed in 306 medical diagnostic radiology centers in 14 provinces in Syria. We employed an X-ray beam analyzer device (NERO model 8000, Victoreen, USA), which was tested and calibrated at the National Secondary Standard Dosimetry Laboratory traceable to the IAEA Network of Secondary Standard Dosimetry Laboratories. Standard QC tool kits were used to evaluate tube and generator of the X-ray machines, which constituted potential (kVp), timer accuracy, radiation output consistency, tube filtration, small and large focal spot sizes, X-ray beam collimation and alignment, as well as high- and low-resolution and entrance surface dose in fluoroscopy. Results: According to our results, most of the assessed operating parameters were in compliance with the standards stipulated by the National Regulatory Authority. In cases of noncompliance for the assessed parameters, maximum value (28.77%) pertained to accuracy of kVp calibration for radiography units, while the lowest value (2.42%) belonged to entrance surface dose in fluoroscopy systems. Conclusion: Effective QC program in diagnostic radiology leads to obtaining information regarding quality of radiology devices used for medical diagnosis and minimizing the doses received by patients and medical personnel. The findings of this QC program, as the main part of QA program, illustrated that most of the considered diagnostic X-ray devices had acceptable performance and few of them need to be recalibrated for some parameters.
    کلید واژگان
    Diagnostic X-Ray
    Fluoroscopy
    Quality Control
    Radiography
    Medical Physics
    Physics of Diagnostics Radiology
    Radiation Detection & Measurement

    شماره نشریه
    2
    تاریخ نشر
    2017-06-01
    1396-03-11
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Atomic Energy Commission of Syria
    Atomic Energy Commission of Syria
    Atomic Energy Commission of Syria

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2017.19712.1186
    http://ijmp.mums.ac.ir/article_8489.html
    https://iranjournals.nlai.ir/handle/123456789/324764

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب