    Development of Computed Tomography Head and Body Phantom for Organ Dosimetry

    (ندگان)پدیدآور
    Akpochafor, MichaelADENEYE, SAMUELOlolade, KehindeOMOJOLA, AKINTAYOAjibade, OLUWAFEMIAdedewe, NusiratAdedokun, AderonkeAweda, MosesAboyewa, Oluyemi
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    1.059 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: Quality assurance in Computed tomography (CT) centers in developing countries are largely hindered by the unavailability of CT phantoms. The development of a local CT phantom for the measurement of organ radiation absorbed dose is therefore requisite. Material and Methods: Local CT phantoms were designed to meet the standard criteria of 32 cm diameter for body, 16 cm diameter for head, and 14 cm in length respectively. The outer plastic shell was made using poly (methyl methacrylate [PMMA]) sheet. The developed CT phantoms were validated against a standard phantom. Radiation absorbed dose was determined by scanning the setup with the same protocol used for the standard phantom. The local phantoms were then verified for organ radiation absorbed dose measurement using bovine tissues. The set up was CT-scanned, and Hounsfield units (HU) for bovine tissues were obtained. Results: There was no significant difference between the local and standard head phantoms (P=0.060). Similarly, no difference was noted between the local and standard body phantoms (P=0.795). The percentage difference in volume CT dose index (CTDIvol) between the body (local and standard) phantoms was higher than that for the head phantoms. There were no significant differences in HU between bovine and human brain, liver, kidney and lung tissues (P=0.938). Conclusion: The local phantoms showed good agreement with the standard ones. The developed phantoms can be used for CT organ radiation absorbed dose measurement in radiology departments in Nigeria.
    کلید واژگان
    Phantom
    Organ
    Radiation Dosage
    Computed Tomography
    Thermoluminescent Dosimetry
    Computed Tomography (CT)
    Medical Physics
    Radiation Detection & Measurement

    شماره نشریه
    1
    تاریخ نشر
    2019-01-01
    1397-10-11
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Department of Radiation Biology, Radiotherapy and Radiodiagnosis
    Department of Radiation Biology, Radiotherapy, and Radiodiagnosis, College of Medicine, Lagos University Teaching Hospital, Idi-Araba, Lagos, Nigeria
    Department of Radiation Biology and Radiotherapy, College of Medicine, Lagos University Teaching Hospital, Idi-Araba.
    FMC ASABA DELTA STATE
    Department of Radiation Biology and Radiotherapy, College of Medicine, University of Lagos Idi-Araba.
    Department of Radiation Biology and Radiotherapy, College of Medicine, University of Lagos.
    Department of Radiation Biology and Radiotherapy, College of Medicine, University of Lagos.
    Department of Radiation Biology, Radiotherapy, Radiodiagnosis and Radiography, College of Medicine, Lagos University Teaching Hospital, Idi-Araba, Lagos, Nigeria
    Department of Physics, Federal University of Technology, Akure, Ondo State, Nigeria

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2018.30906.1360
    http://ijmp.mums.ac.ir/article_10750.html
    https://iranjournals.nlai.ir/handle/123456789/324279

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب