    Quality Control Assessment of Conventional Radiology Devices in Iran

    (ندگان)پدیدآور
    Asadinezhad, MohsenBahreyni Toossi, Mohammad TaghiEbrahiminia, AliGiahi, Masoumeh
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    449.9کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction Quality control (QC) techniques are used in monitoring and maintenance of the components of an x-ray system. QC of radiology devices plays a significant role in reduction of medication dose and optimization of image quality. This study aimed to conduct QC tests on randomly selected radiology devices, installed in diagnostic imaging departments of Iran. Materials and Methods In total, quality control tests were conducted on 51 conventional radiology devices installed in 20 cities of Iran in order to assess the accuracy of peak kilovoltage  (kVp), exposure time, exposure linearity and reciprocity, reproducibility of exposure and determination of half-value layer (HVL) using a calibrated Mult-O-Meter. Results In this study, 38.6% of devices had intolerable variance of kVp accuracy. The results of 34.5% of devices were out of the acceptable limits in exposure time accuracy test. In 46.7% and 53.1% of devices, variance was greater than the acceptable range for exposure linearity and exposure reciprocity, respectively. In terms of reproducibility of exposure test, the reproducibility variance and percentage of tube output variations in 19.4% of devices exceeded the limits. Moreover, the thickness of first HVL was lower than the acceptable limit in 14.7% of devices. Conclusion According to the results of this study, there were wide variations in QC test results, perhaps mainly due to the fact that it is not an obligation to implement QC programs in Iran. The most important problems were non-reciprocity of exposure, nonlinearity of exposure with milliampere-second (mAs), kVp and timer inaccuracy. Involvement of medical physicists, radiologists and radiographers in the implementation of QC programs at various stages of development, installation and use of equipment should enable the gradual improvement in equipment performance.
    کلید واژگان
    Quality Control
    Diagnostic X-Ray
    Radiology
    Medical Physics
    Physics of Diagnostics Radiology

    شماره نشریه
    1
    تاریخ نشر
    2017-03-01
    1395-12-11
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Department of Radiology Technology, School of Paramedical Sciences, Mashhad University of Medical Sciences
    Professor of Medical Physics, Mashhad University of Medical Sciences, Mashhad, Iran
    Assistant professor of medical physics, Biochemistry and Medical Physics Dept., School of Medicine, Guilan University of Medical Sciences, Rasht, Iran
    Student of Radiology Technology, Department of Radiology Technology, School of Paramedical Sciences, Mashhad University of Medical Sciences, Mashhad, Iran

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2017.19052.1173
    http://ijmp.mums.ac.ir/article_8402.html
    https://iranjournals.nlai.ir/handle/123456789/324262

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب