    Design and Fabrication of the Mechanical Part of a Prototype Multileaf Collimator System

    (ندگان)پدیدآور
    Bahreyni Toosi, Mohammad TaghiHashemian, AbdorezaNasseri, ShahrokhRafati Hemayati, Hassan
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    2.992 مگابایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: Multileaf collimator (MLC) is state of the art among the radiation field shaping systems for conformal radiotherapy techniques. Some advantages of this system are as follow: reduction of treatment time, higher precision in conforming the radiation field to the projection of the target volume and lowering the chance of errors made by technologists. The purpose of this research was to design and construct the mechanical section of a prototype MLC.  Material and Methods: The MLC consists of a base plate, 52 leaves, 2 leaf carriages, 2 nuts and 52 triangular thread screws. The leaves are arranged in two opposing banks each containing 26 leaves. The carriages are used to extend the movement of the leaves. The components of the MLC are fabricated by machining processes. The MLC is designed to be mounted on the conventional linear accelerators as an add-on accessory. The leaves were made of aluminum. The dimensions of the leaves are 5 x 200 x 50 mm.  Result: The collimator system weighs approximately 37kg excluding the leaves. The projected width of each leaf on the isocenter of a Varian linear accelerator is 10 mm. The movement of the leaves is accomplished by using the triangular thread screws. The screws required to have a low axial backlash which in turn results in lower precision positioning of the leaves. The positioning accuracy of each leaf is approximately 1.4 mm. Discussion and Conclusion: Mashhad MLC system, in it's present form and design weighs 37 kg. but if Aluminum blades are replaced by tungsten ones then the over all weights of MLC may exceed the weight limit a linac may hold. Therefore futher modifications are required to reduce the MLC total weight. It is desirable to use ball screws and roller cages instead of triangular thread screws to achive a higher positioning pereision of the blades
    کلید واژگان
    Multileaf Collimator
    Design
    Fabrication
    Mechanical part
    Medical Physics
    Physics of Radiotherapy

    شماره نشریه
    34
    تاریخ نشر
    2007-12-01
    1386-09-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Professor, Medical Physics Research Center, Bu-Ali Research Institute, Mashhad University of Medical Sciences, Mashhad, Iran
    Assistant Professor, Medical Physics Research Center, Bu-Ali Research Institute, Mashhad University of Medical Sciences, Mashhad, Iran
    Ph.D. Student in Medical Physics, Mashhad University of Medical Sciences, Mashhad, Iran
    M.Sc. in Industrial Design, Pouian Medical Equipment Engineering Company, Mashhad, Iran

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2007.7547
    http://ijmp.mums.ac.ir/article_7547.html
    https://iranjournals.nlai.ir/handle/123456789/324231

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب