    Design and Implementation of a Complementary Treatment Planning Software for the GZP6 HDR Brachytherapy System (GZP6 CTPS)

    (ندگان)پدیدآور
    Hariri Tabrizi, SanazKamali Asl, AlirezaAzma, Zohreh
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    882.6کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: Brachytherapy is one of the most common treatment modalities for gynecological cancer. The GZP6 brachytherapy system is one of the devices utilized in Iran. It has been considered particularly due to its low cost compared to other more complete and established systems. This system has some deficiencies including lack of a treatment planning software for non-predefined treatments, inability to change the gradually changeable dosimetric variables and using a point source estimation in dose calculation. This report presents a complementary treatment planning software (CTPS) to the system's own dedicated program. Material and Methods: First, the dosimetric characteristics of three GZP6 sources were calculated based on the TG-43 protocol using the MCNP4C Monte Carlo code. Then, the calculated dose distribution around the implanted applicators, based on the selected dwell positions and dwell times, was shown in a graphical user interface (GUI) written using the MATLAB software. Results: The computation uncertainty in the resulting TG-43 parameters was about 1% and the calculated parameters were in good agreement with similar studies on cobalt-60 source dosimetry. Furthermore, the GUI is prepared as a user-friendly executable file which can be installed on any operating system. Discussion and Conclusion: Since different patients have distinct anatomy and physical conditions, a program for non-predefined situations of source arrangement is necessary. Using GZP6 CTPS can satisfy this requirement.
    کلید واژگان
    Brachytherapy
    GZP6 System
    Treatment Planning Software
    TG-43
    Medical Application of Computer Simulation
    Medical Physics
    Physics of Radiotherapy

    شماره نشریه
    1
    تاریخ نشر
    2011-03-01
    1389-12-10
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Ph.D. Student of Radiation Medicine Engineering, , Shahid Beheshti University, Tehran, Iran
    Assistant Professor,Radiation Medicine Engineering., Shahid Beheshti University, Tehran, Iran
    Ph.D. Student of Radiation Medicine Engineering, , Shahid Beheshti University, Tehran, Iran

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2011.7203
    http://ijmp.mums.ac.ir/article_7203.html
    https://iranjournals.nlai.ir/handle/123456789/324141