    Quality Control of Radiography Equipments in Golestan Province of IRAN

    (ندگان)پدیدآور
    Khoshbin Khoshnazar, AlirezaHejazi, PeymanMokhtarian, MiladNooshi, Shahin
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    358.8کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Paper
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction The main goal of this study was to perform quality control test on all radiography units operating in Golestan province of IRAN. Materials and Methods Forty-four X-ray units were examined based on general accepted programs for quality control. Eight parameters including kVp accuracy, kVp reproducibility, mA-time reciprocity, exposure linearity, exposure reproducibility, timer accuracy, filtration, and beam alignment were measured and calculated. Measurements were carried out by a Baracuda X-ray beam analyzer. Results Variance of kVp reproducibility was acceptable in 100% of equipments. kVp accuracy was found to be unsatisfactory in 29.5% of equipments. Variance of mA-time reciprocity was measured to be within reliable limits. Thirty-nine percent of radiography equipments showed non-linear exposure attitude while 16.7 % of them exhibited unacceptable reproducibility of exposure. Moreover, beam misalignment was met in 29.5% of equipments. In 43.2% of radiography equipments, timer accuracy was out of permissible range. Conclusion Timer inaccuracy seems to be a common problem for X-ray units. Exposure non-linearity, mA-time non-reciprocity, kVp inaccuracy, beam misalignment, and finally non-reproducibility of exposure were found to have less importance.
    کلید واژگان
    Golestan Province
    Quality Control
    Radiography Equipment
    Medical Physics
    Physics of Diagnostics Radiology

    شماره نشریه
    1
    تاریخ نشر
    2013-03-01
    1391-12-11
    ناشر
    Mashhad University of Medical Sciences
    سازمان پدید آورنده
    Medical Physicist, Department of Biochemistry and Medical Physics, Faculty of Medicine, Golestan University of Medical Sciences.
    Medical Physicist, Department of Pharmacology and Medical Physics, Faculty of Medicine, Semnan University of Medical sciences
    Radiology Technologist, 5th Azar Hospital, Golestan University of Medical Sciences.
    Health Engineer, Health Affair of Golestan University of Medical Sciences, Golestan University of Medical Sciences

    شاپا
    2345-3672
    URI
    https://dx.doi.org/10.22038/ijmp.2013.917
    http://ijmp.mums.ac.ir/article_917.html
    https://iranjournals.nlai.ir/handle/123456789/324119

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب