    Comparison between Clinically Used Irregular Fields Shielded by Cerrobend and Standard Lead Blocks

    (ندگان)پدیدآور
    Farajollahi, A RBouzarjomehri, FKiani, M
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    425.0کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Research
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Introduction: In radiation therapy centers across Iran, protection of normal tissues is usually accomplished by either Cerrobend or lead block shielding. In this study, the influence of these two shielding methods on central axis dose distribution of photon beam a Cobalt unit was investigated in clinical conditions. Materials and Methods: All measurements were performed for 60Co γ-ray beams and the Cerrobend blocks were fabricated by commercial Cerrobend materials. Standard lead block shields belonged to Cobalt unit. Data was collected through a calibrated ionization chamber, relative dosimetry systemsand a TLD dosimetery.Results: Results of the percent depth dose (PDD) measurements at depths of 0.5, 1, 5, 10, 15 and 20 cm for 23 different field sizes of patients with head and neck cancer showed no significant differences between lead and Cerrobend shielding methods. Measurement results of absolute dosimetry in depths of 1.5, 3, 5, 7, 10 and 12 cm also showed no significant differences between these two shielding methods. The same results were obtained by TLD dosimetry on patient skin.  Conclusion: Use of melt shielding methods is a very easy and fast shield-making technique with no differences in PDD, absolute and skin dose between lead and Cerrobend block shielding methods.
    کلید واژگان
    Shield
    Cerrobend
    Lead block
    Dosimetry
    Radiation Therapy
    TLD dosimetry

    شماره نشریه
    2
    تاریخ نشر
    2015-06-01
    1394-03-11
    ناشر
    Shiraz University of Medical Sciences

    شاپا
    2251-7200
    URI
    https://jbpe.sums.ac.ir/article_43191.html
    https://iranjournals.nlai.ir/handle/123456789/26645

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب