    Rectal and Bladder Dose Measurements in the Intracavitary Applications of Cervical Cancer Treatment with HDR Afterloading System: Comparison of TPS Data with MOSFET Detector

    (ندگان)پدیدآور
    Singh, NAhamed, SSinha, ASrivastava, SPainuly, N KMandal, APrasad, S N
    Thumbnail
    دریافت مدرک مشاهده
    FullText
    اندازه فایل: 
    711.1کیلوبایت
    نوع فايل (MIME): 
    PDF
    نوع مدرک
    Text
    Original Research
    زبان مدرک
    English
    نمایش کامل رکورد
    چکیده
    Background: Intracavitary brachytherapy plays a major role in management of cervical carcinoma. Assessment of dose received by OAR's therefore becomes crucial for the estimation of radiation toxicities in HDR brachytherapy. Objective: Purpose of this study is to evaluate the role of in vivo dosimetry in HDR brachytherapy and to compare actual doses delivered to OAR' s with those calculated during treatment planning. Material and Methods: In this retrospective study, 50 patients of cervical carcinoma were treated by Microselectron HDR. Out of 50 patients, 26 were treated with a dose of 7 Gy and 24 with a dose of 9 Gy, prescribed to point A. Brachytherapy planning and evaluation of dose to bladder and rectum was done on TPS & in vivo dosimetry was performed using portable MOSFET. Results: Calibration factors calculated for both dosimeters are almost equal and are 0.984 cGy/mV and 1.0895 cGy/mV. For bladder, dose deviation was found to be within ± 5% in 28 patients, ± 5-10% in 14 patients, ± 10-15% in 4 patients. Deviation between TPS-calculated dose and dose measured by MOSFET for rectum was within ± 5% in 31 patients, ± 5–10% in 8 patients, and ± 10–15% in 7 patients. Conclusion: TPS calculated doses were slightly higher than that measured by MOSFET. The use of small size MOSFET dosimeter is an efficient method for accurately measuring doses in high-dose gradient fields typically seen in brachytherapy. Therefore, to reduce the risk of large errors in dose delivery, in vivo dosimetry can be done in addition to TPS computations.
    کلید واژگان
    MOSFET
    ICRU
    TPS
    Brachytherapy
    In Vivo Dosimetry
    Radiation Dosimeters

    شماره نشریه
    2
    تاریخ نشر
    2020-04-01
    1399-01-13
    ناشر
    Shiraz University of Medical Sciences
    سازمان پدید آورنده
    PhD, Department of Radiotherapy, King George Medical University, Lucknow 226003, Uttar Pradesh, India
    MSc, Department of Radiation Physics, MNJ Institute of Oncology and Re­gional Cancer Centre, Hyderabad, 500004, Telanga­na
    PhD, Department of Radiotherapy, J.K.Cancer Institute, Kanpur 208002, Uttar Pradesh, India
    MSc, Department of Radiotherapy, King George Medical University, Lucknow 226003, Uttar Pradesh, India
    PhD, Department of Radiotherapy, King George Medical University, Lucknow 226003, Uttar Pradesh, India
    PhD, Department of Radiotherapy and Radiation Medicine, Institute of Medical Sciences, Banaras Hindu University, Varanasi, 221005, India
    MD, Department of Radiotherapy, J.K.Cancer Institute, Kanpur 208002, Uttar Pradesh, India

    شاپا
    2251-7200
    URI
    https://dx.doi.org/10.31661/jbpe.v0i0.1065
    https://jbpe.sums.ac.ir/article_45707.html
    https://iranjournals.nlai.ir/handle/123456789/26384

    مرور

    همه جای سامانهپایگاه‌ها و مجموعه‌ها بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌هااین مجموعه بر اساس تاریخ انتشارپدیدآورانعناوینموضوع‌‌ها

    حساب من

    ورود به سامانهثبت نام

    آمار

    مشاهده آمار استفاده

    تازه ترین ها

    تازه ترین مدارک
    © کليه حقوق اين سامانه برای سازمان اسناد و کتابخانه ملی ایران محفوظ است
    تماس با ما | ارسال بازخورد
    قدرت یافته توسطسیناوب