Chromium(III) ion selective electrode based on 2H-1,4-benzothioazine-2,3(4H)dione dioxime as a neutral carrier
(ندگان)پدیدآور
Fekri, M.H.Darvishpour, M.Baghdar, E.نوع مدرک
TextResearch Paper
زبان مدرک
Englishچکیده
A poly(vinyl chloride) membrane based on a new a-dioxime derivative as membrane carrier was prepared andinvestigated as a Cr(III) selective electrode. The electrode exhibits a good potentiometric response for Cr(III)over a wide concentration range 1.0 x 10-6 to 1.0 x 10-1 M with a slope 19.5+0.5 mV/decade and low detectionlimit of 8.9 x 10-7 M. It has a fast response time having the PVC-ligand-acetophenone-oleic acid composition 6:1:13:1. The proposed electrode works well in awide p1-1 range 1.5-5.5. The proposed sensor was successfuly used for the determination of Cr(III) ions insamples and real samples and as indicator electrode in potentiometric titration of Cr(III) ion. The application ofthis electrode for Cr(III) determination in real samples is described. The results obtained with this procedure arein close agreement with those obtained using AA spectrophotometry.
کلید واژگان
Ion selective electrodePVC membrane
Cr(III) determination
a-dioxime
شماره نشریه
1تاریخ نشر
2011-05-011390-02-11
ناشر
Tehran, Islamic Azad University of Iran, Science and Research Branchسازمان پدید آورنده
--
-