نمایش مختصر رکورد

dc.contributor.authorResaneh, samiraen_US
dc.contributor.authorRajabi, hoseinen_US
dc.contributor.authorBitavafan-Rahabi, ahmaden_US
dc.contributor.authorFarsi-nejad, mohammad rezaen_US
dc.date.accessioned1399-08-21T21:43:32Zfa_IR
dc.date.accessioned2020-11-11T21:43:32Z
dc.date.available1399-08-21T21:43:32Zfa_IR
dc.date.available2020-11-11T21:43:32Z
dc.date.issued2006-06-01en_US
dc.date.issued1385-03-11fa_IR
dc.identifier.citationResaneh, samira, Rajabi, hosein, Bitavafan-Rahabi, ahmad, Farsi-nejad, mohammad reza. (2006). The effect of earth magnetic field on the function of nuclear medicine imaging systems (SPECT). scientific magazine yafte, 8(1), 39-43.en_US
dc.identifier.issn1563-0773
dc.identifier.urihttp://yafte.lums.ac.ir/article-1-1050-en.html
dc.identifier.urihttps://iranjournals.nlai.ir/handle/123456789/483250
dc.description.abstractBackground: The nuclear medicine systems are very sensitive to the variation of the magnetic field. Photomultiplier tubes amplify low energy light signal and change it into electrical current. Low magnetic field can produce some variations in uniformity and special resolution of SPECT systems. In this study, we tried to evaluate the effects of magnetic field on the function of nuclear medicine imaging system, when system rotates in different angles. Materials and Methods: In this study, three SPECT systems with different orientation were examined. A point and line source was placed in the front of the system and when the gantry system rotated in earth magnetic field, variation of energy photo peak, spatial resolution and linearity were determined in different angles. These measurements repeated when an air conditioner was placed at 2-meter distance of the camera gantries. Results were compared using t-student test. Findings: The t-test results show that in all three SPECT systems, the variations of energy photo peak, spatial resolution and linearity were statistically significant (p<0.01). The surrounding magnetic field can effect on accurate function of nuclear medicine systems. Conclusion: The surrounding magnetic field can effect on accurate function of nuclear medicine systems.en_US
dc.format.extent278
dc.format.mimetypeapplication/pdf
dc.languageEnglish
dc.language.isoen_US
dc.relation.ispartofscientific magazine yafteen_US
dc.relation.ispartofمجله علمی پژوهشی یافتهfa_IR
dc.subjectMagnetic fielden_US
dc.subjectSPECTen_US
dc.subjectSpecial resolutionen_US
dc.subjectEnergy resolutionen_US
dc.subjectPhoto peaken_US
dc.subjectlinearityen_US
dc.titleThe effect of earth magnetic field on the function of nuclear medicine imaging systems (SPECT)en_US
dc.typeTexten_US
dc.typeResearchen_US
dc.citation.volume8
dc.citation.issue1
dc.citation.spage39
dc.citation.epage43